BioPro HIC HT 疏水性作用層析管柱

BioPro HIC HT is a hydrophobic interaction chromatography column based on a non-porous hydrophilic polymer. It is suitable for analysis of biopharmaceuticals such as antibody-drug conjugates (ADC’s). Optimized surface modification and column packing condition realize high pressure tolerance and high separation ability. Therefore, high-throughput analysis with a faster flow rate can be achieved.

SPECIFICATION
Matrix Hydrophilic non-porous polymer
Particle size 2.3 µm
Bonded phase Butyl group
Usable temp. range 10-60°C
Usable pH range 2-12
Pressure limit 40 MPa

Features

As BioPro HIC HT is designed for the analysis of biopharmaceuticals, it is the first choice for ADCs or MAbs. It provides high resolution with very fast runs and therefore high throughput.

BioPro HIC HT consists of the fine non-porous polymer particle designed for separation of proteins, and has high pressure tolerance with excellent chromatographic resolution by optimized packing condition. This is capable of the rapid analysis without losing the resolution at a high pressure, which is too high to use with the other commercially available HIC column.

Eluent A) 20 mM NaH2PO4-Na2HPO4 (pH 7.0) containing 1.0 M (NH4)2SO4
B) 20 mM NaH2PO4-Na2HPO4 (pH 7.0)/2-propanol (85/15)
Temperature 25°C
Detection UV at 280 nm
Injection 10 µL
Sample Brentuximab vedotin (2.5 mg/mL)

The DAR analysis by using BioPro HIC HT and the competitor columns under two mobile phase conditions varying 2-propanol content was carried out. The competitor A column having high hydrophobicity did not elute the peak #5 completely with the mobile phase containing 2-propanol lower than 15%. The competitor B column eluted all peaks with the 10% 2-propanol mobile phase condition, but it showed the poor separation between the peaks #4 and #5. On the other hand, BioPro HIC HT eluted all peaks completely with the good separation, and another peak was partially separated on the peak #3. Additionally, it is noteworthy that the same DAR values were observed at any content of 2-propanol.

Column 100 X 4.6 mmI.D.
Eluent A) 20 mM NaH2PO4-Na2HPO4 (pH 7.0) containing 1.0 M (NH4)2SO4
B) 20 mM NaH2PO4-Na2HPO4 (pH 7.0)/2-propanol (90/10) or (85/15)
0-100%B (0-15 min), 100%B (15-20 min), 0%B (20-35 min)
Flow rate 0.5 mL/min
Temperature 25°C
Detection UV at 280 nm
Injection 10 µL
Sample Brentuximab vedotin (2.5 mg/mL)

BioPro HIC HT exhibits an excellent batch-to-batch reproducibility. This is suitable for quality control analysis of biopharmaceuticals such as MAb’s.

Column BioPro HIC HT 2.3 μm, 100 X 4.6 mmI.D.
Eluent A) 100 mM NaH2PO4-Na2HPO4 (pH 7.0) containing 2.0 M (NH4)2SO4
B) 100 mM NaH2PO4-Na2HPO4 (pH 7.0)
0%B (0-1 min), 0-100%B (1-11 min), 100%B (11-15 min)
Flow rate 0.5 mL/min
Temperature 25°C
Detection UV at 280 nm
Injection 15 µL
Sample 1. Adalimumab (0.5 mg/mL)
2. Trastuzumab (0.5 mg/mL)
3. Bevacizumab (0.5 mg/mL)