CHIRAL ART Coated type 手性分離管柱(不對稱/鏡相)
CHIRAL ART Coated type silica packing material is coated with polysaccharide derivatives. CHIRAL ART are suitable for the separation of a wide range of chiral compounds, cis-trans isomers, and geometric isomers. CHIRAL ART are available in HPLC columns and in preparative grades, in large (multi kg) quantities.
SPECIFICATION
Base | : organic/inorganic hybrid silica |
Particle size | : 1.9 μm, 3 µm, 5 µm |
Pore size | : 120 Å |
Carbon contents | : 15% |
pH range | : 1-8 |
Lineup
Column/Packing material | Particle size (µm) | Chiral selector | USP Code | |
---|---|---|---|---|
Amylose-C Amylose-C Neo |
3 5 10 20 |
L51 | ||
Cellulose-C | L40 | |||
Usable mobile phase | n-hexane, n-heptane, ethanol, 2-propanol, acetonitrile, etc. |
Amylose-C Neo is a grade-up column/packing material of Amylose-C.
Features
CHIRAL ART provide good peak shapes on ionic and metal coordination compounds.
Column | 5 μm, 250 X 4.6 mmI.D. |
---|---|
Eluent | n-hexane/2-propanol/diethylamine (80/20/0.1) |
Flow rate | 1.0 mL/min |
Temperature | 25°C |
Detection | UV at 230 nm |
Injection | 10 μL (0.1 mg/mL) |
CHIRAL ART have outstanding packed bed stability provided by the use of high-strength super-wide-pore silica and innovative packing technology. The column efficiency and selectivity are maintained, even after sequential gradient tests at a high flow rate (three times higher than the normal flow rate) and a high pressure (rapid pressure change).
CHIRAL ART are suitable for the case where one intends to shorten an analysis time, (re-)equilibration time, and/or column cleaning time by increasing the flow rate. This feature is also effective when a highly viscous solvent needs to be used as a mobile phase on immobilized type columns.
Sequential gradient test | |
Column | 5 μm, 250 X 4.6 mmI.D. |
---|---|
Eluent | A) n-hexane, B) ethanol 0-100%B (0-15 min) |
Flow rate | 3.0 mL/min |
Pressure | 10-30 MPa/run |
Temperature | 37℃ |
Column performance test | |
Column | 5 μm, 250 X 4.6 mmI.D. |
---|---|
Eluent | n-hexane/ethanol (90/10) |
Flow rate | 1.0 mL/min |
Temperature | 37℃ |
Detection | UV at 230 nm |
Sample | trans-Stilbene oxide |
On both analytical and preparative separation, separation behavior of CHIRAL ART is equivalent to that of the competitor’s product.
trans-Stilbene oxide
Column | 5 μm, 250 X 4.6 mmI.D. |
---|---|
Eluent | n-hexane/ethanol (90/10) |
Flow rate | 1.0 mL/min |
Ibuprofen enantiomers were purified by utilizing the recycling mode of a multiple preparative HPLC system, LC-Forte/R. Recycling chromatography is effective when method optimization for chiral isolation is difficult using single column separation. The recycling method offers high-purity and high-recovery purification. Furthermore, no solvent is consumed during recycling, which greatly contributes to the reduction of solvent consumption during purification.
Ibuprofen
Column | CHIRAL ART Cellulose-C 10 µm, 250 X 20 mmI.D. |
---|---|
Flow rate | 20 mL/min |
Detection | UV at 265 nm |
Injection | 5 mL (10 mg/mL) |
Single Column | Recycling | |
Enantiomeric purity (%ee ) Fr.1 (R)(- -)-Ibuprofen Fr.2 (S)(- +)-Ibuprofen |
95.0 96.8 |
98.4 99.2 |
Yield (%) | 84 | 95 |
Solvent consumption (mL solvent/g product) |
9,523 | 1,276 |