YMC-Pack Polyamine II 層析管柱
YMC-Pack Polyamine II is a silica-based packing bonded with polyamine. It is particularly useful for separation of sugars. The column life of YMC-Pack Polyamine II in aqueous mobile phase is longer than conventional silica-based amino columns, and thus is applicable to separation of oligosaccharides using mobile phase with relatively higher water content. In addition, YMC-Pack Polyamine II can be used to separate ionic compounds with a combination of normal-phase mode and weak anion exchange mode.
SPECIFICATION
| Particle size | : 5µm |
|---|---|
| Pore size | : 120Å |
| Usable pH range | : 2.0-7.5 |
POINT
■ Silica gel chemically bonded to polyamine ■ The most suitable column for separation of sugars ■ Solves durability problem of conventional silica-based amino columns ■ Useful for separation of hydrophilic compounds including vitamins ■ Useful for separation of fat-soluble compounds using nonaqueous mobile phase
Features
Even in a hard flow durability test using mobile phase that consists only of water,the retention time of sugars changes little.


1. Fructose 2. Glucose 3. Sucrose 4. Maltose 5. Lactose

YMC-Pack Polyamine II is applicable for separation of fat-soluble vitamins and water-soluble vitamins as a normal-phase column that can be used with water or buffer and various organic solvents.
Tocopherol isomers

| Column | : YMC-Pack Polyamine II 250 x 4.6mmI.D. |
|---|---|
| Eluent | : n-hexane /ethyl acetate(70/30) |
| Flow Rate | : 1.0mL/min |
| Temperature | : 30ºC |
| Detection | : UV at 295nm |
L-ascorbic acid and erythorbic acid

| Column | : YMC-Pack Polyamine II 250 x 4.6mmI.D. |
|---|---|
| Eluent | : 50mM NH4H2PO4 /acetonitrile(30/70) |
| Flow Rate | : 1.0mL/min |
| Temperature | : 30ºC |
| Detection | : UV at 250nm |